| 9132-50 | 9010-50 |
Rated current | AC 20/50/100/200/500/1000 A | AC 10/20/50/100/200/500 A |
Accuracy | ¡¾3% rdg. ¡¾0.2% f.s. (45 to 66Hz) | ¡¾2% rdg. ¡¾1% f.s. (45 to 66Hz) |
Frequency characteristics | Add to amplitude accuracy for frequencies from 40 to 1kHz: ¡¾ 1% rdg. | Add to amplitude accuracy for frequencies from 40 to 1kHz: ¡¾6% rdg. (at 10A and 20A range) ¡¾3% rdg. (for 50A range and above) |
Output rate | 0.2V AC f.s. (f.s.=setting rage) (Connect to a voltage input device providing a high input impedance of 1M§Ù) |
Max. current input | (For 40Hz to500 Hz: 100%, and for 500Hz to 1kHz: within 90% of derating) 1000Arms continuous (all ranges) | (For 40Hz to 100Hz: 100%, and for 100Hz to 1kHz: within 50% of derating) 150Arms continuous (10/20/50A ranges) 400Arms continuous (100/200A ranges) 650Arms continuous (500A range) |
Max. circuit voltage | 600Vrms (Insulated conductor) | 600Vrms (Insulated conductor) |
Diameter | ¨ª 55mm(2.17in), or 20mm(0.79in) ¡¿ 80mm(3.15in) busbar | ¨ª 46mm(1.81in) |
Dimensions, mass | 100mm(3.94in)W ¡¿ 224mm(8.82in)H ¡¿ 35mm(1.38in)D, 600g(21.2oz), cord length: 3m(9.84ft) | 78mm(3.07in)W ¡¿ 188mm(7.40in)H ¡¿ 35mm(1.38in)D, 420g(14.8oz), cord length: 3m(9.84ft) |